High-temperature Sterilization

Product Code: 150YDA/Horizontal cylindrical pressure steam sterilizer

280200150yda

The steam can’t go through to chamber when the door haven’t close , and the pilot off to inform. You can processing dry operate

When the pressure inner over 0.027Mpa, the door be lock and not be open

If you not choose the sterile time and start machine, buzzer sound

If you mistake open the door in the course of stat, the jacket valve auto off and open the chamber valve

Lack water the pilot on, heating and sterile pilot off, the buzzer sound and auto release steam

Overload and leakage protection, when leakage or over current occur, it auto off the power

Temperature controller not work  and the inner pressure over 0.165Mpa, the safety valve auto start open

Capacity: 150L(D.480mmx850mm)
Working pressure: 0.22MPa
Working temperature: 134℃
Adjustment of temperature: 40-134℃
Timer : 0-60 min

Time for drying: 0-60min

Heat average: ≤± 2 ℃

Power: 380V 50Hz/6Kw

 

 

Product Code: 200YDA/Horizontal cylindrical pressure steam sterilizer

Capacity: 200L(D.515mmx1000mm)

Power: 380V 50Hz/9Kw

 

Product Code: 280YDA/Horizontal cylindrical pressure steam sterilizer

Capacity: 380L(D.600mmx1000mm)

Power: 380V 50Hz/12Kw

Product Code: 400YDA/Horizontal cylindrical pressure steam sterilizer

Capacity: 400L(D.00mmx1100mm)

Power: 380V 50Hz/18Kw

 

Product Code: 500YDA/Horizontal cylindrical pressure steam sterilizer

 

Capacity: 500L(D.700mmx1300mm)

Power: 380V 50Hz/18Kw


Product Code: 150YDBH/orizontal cylindrical pressure steam sterilizer (microcomputer control)

280200150ydb

Sterilizing course automatic sterilization controlled by computer, easy to operate. 

With drying function, suitable dressing drying. 

With over-temperature, over-pressure auto-protect device. 

The door opening mechanism cannot be operated until the pressure in the chamber is reduced to 0.027MPa.It can’t be start on if the doesn’t closed well. 

The safety value will be open automatically when the inner pressure over 0.24MPa, and the steam be exhausting to the water tank. 

The power will be cut-off automatically, cut-off the water and alarm when the machine lacking water.

Capacity: 150L(D.480mmx850mm)
Working pressure: 0.22MPa
Working temperature: 134℃
Adjustment of temperature: 40-134℃
Timer : 0-99 min

Time for drying: 0-99min

Heat average: ≤± 2 ℃

Power: 380V 50Hz/6Kw

 

Product Code: 200YDB/Horizontal cylindrical pressure steam sterilizer (microcomputer control)

Capacity: 200L(D.515mmx1000mm)

Power: 380V 50Hz/9Kw

 

Product Code: 280YDB/Horizontal cylindrical pressure steam sterilizer (microcomputer control)

Capacity: 280L(D.515mmx1000mm)

Power: 380V 50Hz/12Kw


Product Code: 150YDC/Horizontal cylindrical pressure steam sterilizer

280200150ydc

The steam can’t go through to chamber when the door haven’t close , and the pilot off to inform. You can processing dry operate

When the pressure inner over 0.027Mpa, the door be lock and not be open

If you not choose the sterile time and start machine, buzzer sound

If you mistake open the door in the course of stat, the jacket valve auto off and open the chamber valve

Lack water the pilot on, heating and sterile pilot off, the buzzer sound and auto release steam

Overload and leakage protection, when leakage or over current occur, it auto off the power

Temperature controller not work  and the inner pressure over 0.165Mpa, the safety valve auto start open

Capacity: 150L(D.480mmx850mm)
Working pressure: 0.22MPa
Working temperature: 134℃
Adjustment of temperature: 40-134℃
Timer : 0-60 min

Time for drying: 0-60min

Heat average: ≤± 2 ℃

Power: 380V 50Hz/6Kw

 

Product Code: 200YDC/Horizontal cylindrical pressure steam sterilizer

Capacity: 200L(D.515mmx1000mm)

Power: 380V 50Hz/9Kw

 

Product Code: 280YDC/Horizontal cylindrical pressure steam sterilizer

 

Capacity: 280L(D.600mmx1000mm)

Power: 380V 50Hz/12Kw